SNR Value Evaluation of Knee MRI 1.5 T Using Volume and Surface Coil in Fatmawati General Hospital Jakarta

Nursama Heru Apriantoro, Sriyatun, Samsun, Guntur Winarno, Fitri Aryani, Pramono, Dimas Muhammad Avicenna, Nadya Octafiani Poernama

Politeknik Kesehatan Kementerian Kesehatan, Fatmawati General Hospital Jakarta, Indonesia

journal of social and political sciences
pdf download

Download Full-Text Pdf

doi

10.31014/aior.1994.02.01.14

Abstract

Introduction: Signal to noise ratio (SNR) is one of the determining factors of MRI images quality. SNR is defined as the ratio of the intensity of an object signal seen with noise. There are several factors that can affect SNR, one of them is the radiofrequency coil. Knee MRI examination based on the literature using a Knee coil with the shape of the volume. While in some hospitals the use of coil when the MRI knee examination can also use the surface coil. This is because the hospital does not have the coil volume or the condition of the patient's knee that it is not possible to use volume coil. Objective : To evaluate SNR on MRI Knee examination using volume coil and surface coil with 1.5T Magnetom Esenzza MRI Scanners at Fatmawati Hospital. Method: The type of this research is quantitative analytics with experimental method, which was conducted in Radiology installation of Magnetic Resonance Imaging (MRI) atFatmawati General Hospital Jakarta starting from March to April 2018, with 16 samples. The technique of data collecting is done by observation and experiment. Research instrument consists of the worksheet, operator console, and counting sheet. Processing and data analysis is done by using paired sample T-test statistic. Results: The result of this study indicates that there is a difference of image quality in SNR aspect on the image of MRI knee inspection using volume coil and surface coil marked with p-value <0.05. Conclusion: MRI Knee examination using volume coilpossess higher SNR than using the surface coil. But, the surface coil can be the second choice for MRI Knee examination if the hospital doesn't have volume coil or surface coil, it can't be used for the patient.

References

  1. Ihalainen T. Quality control methods for magnetic resonance imaging in a multi-unit medical imaging organization. 2016.

  2. Chrysikopoulos HS. Clinical MR Imaging and Physics. Berlin: Springer; 2009. 171 p.

  3. Brian M. Dale PM, Mark A. Brown P, Richard C. Semelka M. MRI Basic Principles and Applications. Fifth Edit. West Sussex: WILEY Blackwell; 2015. 225 p.

  4. Jackson DW, Jennings LD, Maywood RM, Berger. PE Magnetic resonance imaging of the knee. 1988;16(1):14–6.

  5. Spouge AR, Pope TL. Practical MRI of the Foot and Ankle. Spouge AR, Pope TL, editors. United State: CRC Press; 2001.

  6. Mayahi R, Erturan G, Jackson W, Price A. The Knee: Surface Coil MR Imaging at 1,5 T. Radiology [Internet]. 1986;159(3):747–51. Available from: https://www.scopus.com/inward/record.uri?eid=2-s2.0-84944054780&doi=10.1016%2FB978-0-323-09138-1.00077-2&partnerID=40&md5=3d80b73e9e90813a3a0b31f6c7798d3c

  7. Lawrence D, Kanal E, Brunberg JA, Johnstone GF, Swensen HE, Wolf GL. 1,5 T Surface Coil MRI of the Knee. AJR Am J Roentgenol. 1986;147:293–300.

  8. H J, Levy T, V J. Menisci of the Anterior of the Knee : Cruciate Ligament. Radiology. 1988;(3):769–74.

  9. Westbrook C, Roth CK, Talbot J. MRI IN PRACTICE 4th EDITION. 4th ed. United Kingdom: WILEY Blackwell; 2011. 427 p.

  10. Liney G. MRI From A to Z A Definitive Guide for Medical Professionals. 1st ed. United State: CAMBRIDGE; 2005. 249 p.

  11. Weishaupt D, Köchli VD, Marincek B. How Does MRI Work ? An Introduction to the Physics and Function. 2nd ed. Heilmann DU, editor. Berlin: Springer;

  12. Rohan JW, Yokochi C, Drecoll EL. Color Atlas of Anatomy A Photographic Study of the Human Body. 7th Editio. Germany: Schattauer; 2011. 503 p.

  13. Cooper G. Pocket Guide to Musculoskeletal Diagnosis. Totowa: HUMANA PRESS; 2006.

  14. Brink WM, Webb AG. High Permittivity Pads Reduce Specific Absorption Rate, Improve B 1 Homogeneity, and Increase Contrast-to-Noise Ratio for Functional Cardiac MRI at 3 T. 2014;1640:1632–40.

  15. McRobbie DW. MRI from Picture to Proton [Internet]. 2bd Editio. New York: CAMBRIDGE; 2003. 380 p. Available from: https://books.google.nl/books?id=gfuO6NK_InkC

  16. Blink EJ. BASIC mri : Physics. Mri: Physics, editor. 2004. 0-75 p.